HCA1 Chlorine Plus Photometer System Instruction Manual Ideal for the measurement of Free and Total Chlorine in Drinking Water, Waste Water, Enviro
4. Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes the potential for cross contamination from a previou
HCA1 Micro CL (DPD-1, -4) Strip Interferences Interfering Substance Interference Level and treatment Acidity If sample has acidity above 150mg/L CaCO3
8. Minimize exposure of meter and test reagents to heat above 32°C (90°F). 9. Dry the outside of the meter when testing is complete or before storag
Parts Order/Reorder Information Part# Description 1000040-1 Meter, case and 2 bottles test strips, Free chlorine and total Chlorine 1000040-2 Meter,
Contents HCA1 Chlorine Plus Specifications ... 4 HCA1 Range and Accuracy ...
HCA1 Chlorine Plus Specifications Measurement method: Photometric Light source: Light Emitting Diode (LED) Wavelength: 525 nm Transmissio
HCA1 General Information In order to save power, the meter is designed to turn off after 3 minutes (timed from the last button pressed). Should the me
calibration, the Ready Snap® 3 Calibration Verification Standard part #490903) is available for purchase. Compliance Verification for Free and Total C
2. Remove the used batteries and install 4 new AAA batteries following the diagram for correct polarity (see diagram). We recommend high quality AAA
MENU CL1 DPD-4 Total Chlorine Test Procedure 1. Remove (1) Micro CL (DPD-4) Strip from the bottle before beginning the test. Place it in a dry conven
(in sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample. 7. The cursor will move across the display while